• en
  • de

Convergys® Elisa 96X

The Convergys® EL-Reader 96X can perform ELISA, EIA or enzymatic activity measurements in the wavelength range between 400 and 750 nm using end point, two point, and kinetic measurement methods.

  • Flexible and high performance measurement
  • Multiple measurement modes and analysis methods
  • Wavelength range from 400 to 750 nm
  • Built-in incubator for kinetic studies
  • Built-in shaker with 8 Hz, 11 Hz, and 14 Hz user selectable shaking modes
  • 8x21 character high contrast graphic LCD
  • Convergys® ELR 96X PC software to control measurements and data management through PC
  • Easy-to-use menu with soft key interface
  • Additional Filter cassettes are optionally available
  • Wavelength range: 400–750 nm
  • Interference filters: Max 8 selectable filters in a filter wheel, 10 +/-2 nm half bandwidth filter, center wavelength accuracy +/-2 nm
  • Measurement range: 0.000 – 4.000 Abs
  • Resolution: 0.001 Abs
  • Accuracy: +/- 0.005 Abs or 2 %, whichever is greater (from 0.000 to 3.000 Abs)
  • Precision: <0.3 % CV from 0.000 to 2.000 Abs; <1.0 % CV from 2.000 to 3.000 Abs
  • Linearity: +/- 2 % from 0.000 to 3.000 Abs
  • Measurement time: <5 seconds at continuous reading
  • Measurement mode: End point, two point, and kinetic
  • Light source: Tungsten halogen lamp (12 V, 20 W)
  • Photo detector: 9 silicon photo detectors
  • Shaking: Three user selectable modes(8 Hz, 11 Hz, and 14 Hz)
  • Incubator: Ambient temp +3 to 50 °C
  • Plate type: 96-well plate
  • Display: 8x21 character graphic dot-matrix LCD module
  • Keyboard: 25 key tactile membranes
  • Printer interface: Parallel Centronics
  • Data storage: 100 sets of protocol, 50 sets of sample data and 20 sets of standard curve
  • Computer interface: RS-232 or USB serial interface
  • Dimensions: (WxDxH) 355x345x174 mm
  • Weight: 14 kg
  • Power supply: 100–240 V, 50–60 Hz, 90 W max
  • Operating conditions: 15–35 °C, RH 90 % max
http://convergent-technologies.de/product/convergys-elisa-96x/